ATTAIN ABILITY PLUS
MDR report key: 3992686
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-09734
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 16, 2014
- Report Date
- June 16, 2014
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 4076-52 LEAD IMPLANTED (B)(6) 2013; DTBA1D1 ICD IMPLANTED (B)(6) 2013. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT T-WAVE OVERSENSING WAS OBSERVED DURING LEFT VENTRICULAR (LV) AND BI-V PACING. IN ADDITION, DIAPHRAGMATIC STIMULATION WAS IDENTIFIED DURING THE PATIENT'S VISIT TO THE CLINIC. REPROGRAMMING WAS DONE FOR THE LEFT VENTRICLE AND RIGHT VENTRICLE LEADS. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469460 | ATTAIN ABILITY PLUS | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 429688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00067 YR | Required Intervention | 6935-65 LEAD |