FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3992680 · Received August 8, 2014

Report

Report Number
2649622-2014-09730
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE LEAD EXHIBITED HIGH DEFIBRILLATION THRESHOLDS, BEING UNABLE TO SUCCESSFULLY DEFIBRILLATE THE PATIENT AT MAXIMUM VOLTAGE. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470195 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00026 YR (B)(4) ICD