CONCERTO II CRT-D
Report
- Report Number
- 3004209178-2014-14786
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- February 6, 2014
- Report Date
- May 23, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Removal / Correction Number
- Z-0110-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE ACTUAL LONGEVITY WAS LESS THAN 80% OF THE 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. CONCOMITANT PRODUCT: 4076-52, LEAD, IMPLANTED (B)(6) 2007. (B)(4).
THE DEVICE BATTERY VOLTAGE WAS LOW. THE DEVICE WAS REMOVED AND REPLACED. THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470233 | CONCERTO II CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D274TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00084 YR | Hospitalization| R | 6947-65 LEAD, 4194-78 LEAD |