CAPSURE Z NOVUS
Report
- Report Number
- 2649622-2014-09752
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P850089
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5076-52 LEAD, IMPLANTED (B)(6) 2010. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO. THE OUTER INSULATION OF THE LEAD WAS ALSO OBSERVED TO HAVE BLOOD INGRESSION. VISUAL ANALYSIS FOUND OUTER INSULATION BREACHED IN-VIVO/DEPRESSION. THE BREACHED INSULATION DID CONTRIBUTE TO THE ELECTRICAL COMPLAINT OF HIGH THRESHOLD.
IT WAS REPORTED THAT HIGH THRESHOLD LEVELS WERE OBSERVED ON THE PATIENT'S RIGHT ATRIAL (RA) LEAD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470156 | CAPSURE Z NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5554-45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00058 YR | Hospitalization| R | ADDR01 DEVICE |