FDA Adverse Event Injury Summary report: N

CAPSURE Z NOVUS

MDR report key: 3992637 · Received August 8, 2014

Report

Report Number
2649622-2014-09752
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 23, 2014
Report Date
June 23, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P850089
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5076-52 LEAD, IMPLANTED (B)(6) 2010. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO. THE OUTER INSULATION OF THE LEAD WAS ALSO OBSERVED TO HAVE BLOOD INGRESSION. VISUAL ANALYSIS FOUND OUTER INSULATION BREACHED IN-VIVO/DEPRESSION. THE BREACHED INSULATION DID CONTRIBUTE TO THE ELECTRICAL COMPLAINT OF HIGH THRESHOLD.

Description of Event or Problem · 1

IT WAS REPORTED THAT HIGH THRESHOLD LEVELS WERE OBSERVED ON THE PATIENT'S RIGHT ATRIAL (RA) LEAD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470156 CAPSURE Z NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5554-45

Patients

Seq Age Sex Outcome Treatment
1 00058 YR Hospitalization| R ADDR01 DEVICE