FDA Adverse Event Injury Summary report: N

REVO MRI SURESCAN

MDR report key: 3992598 · Received August 8, 2014

Report

Report Number
3004209178-2014-14796
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWP
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED A FULL ELECTRICAL RESET.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT COMPLAINED OF PACEMAKER SYNDROME AND THE DEVICE WAS FOUND TO HAVE AN ELECTRICAL RESET. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469999 REVO MRI SURESCAN PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MED REL MEDTRONIC PUERTO RICO RVDR01

Patients

Seq Age Sex Outcome Treatment
1 00085 YR Required Intervention 5086MRI X2 LEADS