FDA Adverse Event Injury Summary report: N

KAPPA 700 DR

MDR report key: 3992584 · Received August 8, 2014

Report

Report Number
2647346-2014-00069
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 20, 2014
Report Date
June 20, 2014
Manufacturer
MDT PUERTO RICO OPERATIONS CO, MED REL
Product Code
NVZ
PMA / PMN Number
P980035
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 4092 LEAD, IMPLANTED (B)(6) 2002. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS ALSO REPORTED THAT THE PATIENT HAS HAD SLOW HEART RATES AT NIGHT INTO THE FORTIES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED. THE DEVICE WAS FOUND TO BE AT ELECTIVE REPLACEMENT INDICATOR (ERI). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471480 KAPPA 700 DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MDT PUERTO RICO OPERATIONS CO, MED REL KDR701

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Hospitalization| R 5076 LEAD