FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3992571 · Received August 8, 2014

Report

Report Number
2649622-2014-09791
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5076 LEAD, IMPLANTED: (B)(6) 2011; 1258T LEAD, IMPLANTED: (B)(6) 2011. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS) AND THE ATRIAL LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING DURING AN EPISODE OF ATRIAL TACHYCARDIA. THE LEADS WERE REPROGRAMMED AND REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469935 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 693558

Patients

Seq Age Sex Outcome Treatment
1 00048 YR Required Intervention D314TRG ICD