FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 3992525
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-09818
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- April 6, 2014
- Report Date
- May 6, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE STYLET/GUIDEWIRE WAS KINKED/BUCKLED AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PHYSICIAN HAD DIFFICULTY INSERTING THE STYLET DOWN THE IMPLANTABLE PACING LEAD AFTER PLACEMENT. MULTIPLE STYLETS WERE ATTEMPTED BUT WERE UNABLE TO BE FULLY INSERTED INTO THE LEAD LUMEN. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 472205 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00072 YR | Required Intervention |