FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3992525 · Received August 8, 2014

Report

Report Number
2649622-2014-09818
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
April 6, 2014
Report Date
May 6, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE STYLET/GUIDEWIRE WAS KINKED/BUCKLED AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PHYSICIAN HAD DIFFICULTY INSERTING THE STYLET DOWN THE IMPLANTABLE PACING LEAD AFTER PLACEMENT. MULTIPLE STYLETS WERE ATTEMPTED BUT WERE UNABLE TO BE FULLY INSERTED INTO THE LEAD LUMEN. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472205 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Required Intervention