PROTECTA XT CRT-D
Report
- Report Number
- 3004209178-2014-14818
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 3, 2014
- Report Date
- June 3, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5071-53, LEAD, IMPLANTED: (B)(6) 2011; 5071-53, LEAD, IMPLANTED: (B)(6) 2011. (B)(4).
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 95% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REACHED EARLY ELECTIVE REPLACEMENT INDICATOR (ERI). IT WAS FURTHER REPORTED THAT THE TWO EPICARDIAL LEFT VENTRICULAR (LV) LEADS, WHICH WERE BIFURCATED TO BIPOLAR, HAD HIGH THRESHOLDS. THE LEAD ADAPTER WAS REMOVED AND EACH LEAD WAS TESTED UNIPOLAR, RESULTING IN A LOWER THRESHOLD. ONE EPICARDIAL LEAD WAS CAPPED, AND THE OTHER REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467295 | PROTECTA XT CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D314TRG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00067 YR | 694765, LEAD |