SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2014-09848
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 13, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Removal / Correction Number
- Z-0475-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE DISTAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED. THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. (B)(4).
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC AFTER THE LEAD INTEGRITY ALERT (LIA) TRIGGERED DUE TO NOISE ON THE RIGHT VENTRICULAR (RV) LEAD. A POSSIBLE FRACTURE WAS SUSPECTED. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 468547 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694765 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00075 YR | Hospitalization| R | D284VRC, IMPLANTABLE CARDIOVERTER DEFIBRILLATOR |