FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 3992462 · Received August 8, 2014

Report

Report Number
3004209178-2014-14822
Event Type
Injury
Date Received
August 8, 2014
Report Date
May 21, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NVZ
PMA / PMN Number
P980035
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THAT THE DEVICE HAD AN INTEGRATED CIRCUIT LOCK-UP. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. THIS EVENT WAS ORIGINALLY REPORTED VIA REMEDIAL ACTION EXEMPTION. CONCOMITANT MEDICAL PRODUCTS: 5054-58, IMPLANTABLE PACING LEAD, (B)(6) 2002. (B)(4).

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO EARLY BATTERY DEPLETION. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468080 ENRHYTHM DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MEDTRONIC MED REL MEDTRONIC PUERTO RICO P1501DR

Patients

Seq Age Sex Outcome Treatment
1 00050 YR Hospitalization| R 4068-52, IMPLANTABLE PACING LEAD