VIRTUOSO VR
Report
- Report Number
- 9614453-2014-01978
- Event Type
- Injury
- Date Received
- August 8, 2014
- Report Date
- May 29, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES; HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. THIS EVENT OCCURRED OUTSIDE THE US AND PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. (B)(4).
THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO NO LONGER BEING NEEDED BY THE PATIENT. AN IMPROVEMENT WAS SEEN IN LEFT VENTRICULAR FUNCTION AND NO VENTRICULAR TACHYCARDIA (VT) WAS OBSERVED SINCE A VT ABLATION WAS PERFORMED. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467828 | VIRTUOSO VR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | IPG MFG SWITZERLAND | D164VWC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R | 7022 LEAD |