FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3992440 · Received August 8, 2014

Report

Report Number
2649622-2014-09886
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 4512 LEAD, IMPLANTED: (B)(6) 2003; 4471 LEAD, IMPLANTED: (B)(6) 2003. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED WILDLY VARIABLE HIGH IMPEDANCE WHEN PROGRAMMED TO BIPOLAR. THE PACE/SENSE PORTION OF THE RA LEAD WAS REPROGRAMMED TO UNIPOLAR AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472678 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407645

Patients

Seq Age Sex Outcome Treatment
1 00044 YR Required Intervention C4TR01 CRT-P