FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3992420 · Received August 8, 2014

Report

Report Number
2649622-2014-09896
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH R-WAVES, SO IT WAS DECIDED TO REPOSITION THE LEAD. THE HELIX OF THE RV LEAD WAS RETRACTED AND THE LEAD WAS REPOSITIONED, HOWEVER THE PHYSICIAN WAS UNABLE TO EXTEND THE HELIX OF THE LEAD BACK OUT. THE RV LEAD WAS EXPLANTED AND REPLACED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467713 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M55

Patients

Seq Age Sex Outcome Treatment
1 00078 YR Hospitalization| R DVBB1D4 ICD