SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2014-09896
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 18, 2014
- Report Date
- June 18, 2014
- Manufacturer
- MPRI
- Product Code
- NVY
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH R-WAVES, SO IT WAS DECIDED TO REPOSITION THE LEAD. THE HELIX OF THE RV LEAD WAS RETRACTED AND THE LEAD WAS REPOSITIONED, HOWEVER THE PHYSICIAN WAS UNABLE TO EXTEND THE HELIX OF THE LEAD BACK OUT. THE RV LEAD WAS EXPLANTED AND REPLACED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467713 | SPRINT QUATTRO SECURE | DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES | NVY | MPRI | 6947M55 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00078 YR | Hospitalization| R | DVBB1D4 ICD |