FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3992404 · Received August 8, 2014

Report

Report Number
2649622-2014-09907
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5076-52 LEAD, (B)(6) 2008. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT EXPERIENCED DIZZINESS WHEN THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING. THE RV LEAD WAS REPROGRAMMED AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468003 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00091 YR Required Intervention P1501DR IPG 2008-01-07