FDA Adverse Event
Malfunction
Summary report: N
PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
MDR report key: 3992391
·
Received August 8, 2014
Report
- Report Number
- 2182208-2014-02397
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Report Date
- June 27, 2014
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Removal / Correction Number
- Z-1661-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: DURING ANALYSIS IT WAS DETERMINED THAT THE MAIN PRINTED CIRCUIT BOARD AND INTERCONNECT FLEX WERE OUT OF SPECIFICATION ELECTRICALLY AND INTERMITTENT OPERATION WAS CONFIRMED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. (B)(4).
Description of Event or Problem · 1
THE EXTERNAL PULSE GENERATOR WAS RETURNED AS A LOANER FOR RESTOCK AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Devices