FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 3992308 · Received August 8, 2014

Report

Report Number
2649622-2014-09954
Event Type
Injury
Date Received
August 8, 2014
Date of Event
April 3, 2014
Report Date
May 22, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 6935M55 LEAD, IMPLANTED (B)(6) 2013. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD DISLODGED AND WAS EXPERIENCING LOSS OF CAPTURE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468265 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 419678

Patients

Seq Age Sex Outcome Treatment
1 00086 YR D204TRM CRT-D