FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3992243 · Received August 8, 2014

Report

Report Number
2649622-2014-09988
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 26, 2014
Report Date
June 26, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED, AND THE RV (RIGHT VENTRICULAR) DEFIBRILLATION COIL DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: (B)(4) ICD IMPLANTED: 2009 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPEDANCE ON THE SUPERIOR VENA CAVA (SVC) AND RIGHT VENTRICULAR (RV) LEAD DEFIBRILLATION COILS HAS BEEN HIGH AND ERRATIC AND A FRACTURE IS SUSPECTED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469376 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00083 YR Hospitalization| R 5076-52 LEAD