PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Report
- Report Number
- 2183613-2014-01122
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- April 30, 2014
- Report Date
- June 26, 2014
- Manufacturer
- MEDTRONIC MILACA, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Removal / Correction Number
- Z-1661-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: DURING ANALYSIS IT WAS DETERMINED THAT THE LOWER CASE WAS BROKEN AND THE BATTERY CONTACTS WERE OUT OF SPECIFICATION (MECHANICAL). (B)(4).
THE EXTERNAL PULSE GENERATOR WAS RETURNED AS A LOANER FOR RESTOCK AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
THE EXTERNAL PULSE GENERATOR WAS RETURNED AS A LOANER FOR RESTOCK AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469902 | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC MILACA, INC. | 5348 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |