FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3992169 · Received August 8, 2014

Report

Report Number
2649622-2014-10028
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 11, 2014
Report Date
June 11, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HELIX OF THE RIGHT VENTRICULAR (RV) LEAD APPEARED TO BE OUT DURING IMPLANT. IT WAS GETTING CAUGHT AND CAUSING DIFFICULTY DURING THE PROCEDURE. UPON EXAMINATION THE HELIX APPEARED TO BE BENT. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.