FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3992091 · Received August 8, 2014

Report

Report Number
2183613-2014-01125
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 21, 2014
Report Date
June 30, 2014
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Removal / Correction Number
Z-1661-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE INTERCONNECT FLEX ASSEMBLY WAS OUT OF SPECIFICATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INTERCONNECT COMPONENT FAILED AND WAS UNABLE TO START THE DEVICE. THE DEVICE WAS RETURNED TO THE MANUFACTURER. INFORMATION ABOUT THE EVENT CONTEXT IS UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471204 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MILACA, INC. 5348

Patients

Seq Age Sex Outcome Treatment
1