CONCERTO II CRT-D
Report
- Report Number
- 3004209178-2014-14923
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- November 1, 2013
- Report Date
- March 12, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVENT SUMMARY : THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE INDICATED NORMAL BATTERY DEPLETION.
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE LOST TELEMETRY AND FUNCTION DUE TO BATTERY DEPLETION. THE CAUSE OF THE DEPLETION CANNOT BE DETERMINED. THE DEVICE WAS FULLY FUNCTIONAL AND OPERATED UNDER NORMAL CURRENT DRAIN WHEN POWERED WITH AN EXTERNAL SUPPLY. THE RESIDUAL VOLTAGE AND IMPEDANCE INDICATES THAT THE BATTERY WAS NOT INTERNALLY SHORTED. SINCE NO SAVE TO DISK DATA COULD BE RETRIEVED FROM THE DEVICE AND NO OTHER DATA WAS PROVIDED FROM THE FIELD THERE IS NO WAY TO CONFIRM THIS RESULT. THE RESULT OF ANALYSIS IS INCONCLUSIVE. CONCOMITANT MEDICAL PRODUCTS: 5554-53 LEAD, IMPLANTED (B)(6) 2011; 419688 LEAD, IMPLANTED (B)(6)2011. (B)(4).
THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED AND REPLACED DUE TO NORMAL BATTERY DEPLETION. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470759 | CONCERTO II CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D274TRK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00081 YR | Hospitalization| R | 6944-65 LEAD |