FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3991793 · Received August 8, 2014

Report

Report Number
2649622-2014-07951
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 1, 2014
Report Date
May 2, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS ATTEMPTED TO IMPLANT IN THE RIGHT VENTRICLE (RV) BUT NOT USED. THE LEAD WAS FOUND TO HAVE HIGH IMPEDANCE. ALSO, THE HELIX WOULD EXTEND RAPIDLY AFTER ONLY FOUR TURNS. A NEW LEAD WAS USED TO IMPLANT SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467224 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00087 YR