FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS MRI SURESCAN

MDR report key: 3991758 · Received August 8, 2014

Report

Report Number
2649622-2014-07962
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 1, 2014
Report Date
June 29, 2020
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE DISTAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THAT THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED. THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO ESC (ENVIRONMENTAL STRESS CRACKING) WHILE IN VIVO.

Additional Manufacturer Narrative · 1

NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 7278 ICD, IMPLANTED: (B)(6) 2008. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD HAD AN IMPEDANCE JUMP AND WAS FOUND TO BE HIGH. THE LEAD ALSO HAS A POSSIBLE FRACTURE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE LEAD WAS REMOVED WITH LASER WHEN THE SYSTEM WAS EXPLANTED.

Description of Event or Problem · 1

IT WAS ALSO REPORTED THAT THERE WAS NOISE AND HIGH SHORT INTERVAL COUNTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472182 CAPSUREFIX NOVUS MRI SURESCAN ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407652

Patients

Seq Age Sex Outcome Treatment
1 00046 YR Hospitalization| R 6947 LEAD