FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3991757 · Received August 8, 2014

Report

Report Number
3008973940-2014-00257
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 6, 2014
Report Date
June 7, 2014
Manufacturer
MEDTRONIC SINGAPORE OPERATIONS
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, AFTER THE RIGHT VENTRICULAR (RV) LEAD WAS POSITIONED TESTING REVEALED HIGH IMPEDANCE WITH A POSSIBLE FRACTURE. THE TESTING CABLE WAS CHANGED OUT HOWEVER THE IMPEDANCE PROBLEM PERSISTED. THE RV LEAD WAS REPOSITIONED AND THE PROBLEM STILL PERSISTED. MORE THAN THREE DIFFERENT POSITIONS WERE ATTEMPTED YET THE PROBLEM REMAINED. THE RV LEAD WAS REMOVED, EXCHANGED FOR A DIFFERENT LEAD AND THERE WAS NORMAL THRESHOLD AND IMPEDANCE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467395 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC SINGAPORE OPERATIONS 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00064 YR