FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3991716 · Received August 8, 2014

Report

Report Number
2183613-2014-00976
Event Type
Malfunction
Date Received
August 8, 2014
Report Date
May 22, 2014
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Removal / Correction Number
Z-0430-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THAT ONE BAIL COVER WAS BROKEN, THE RING COVER WAS BROKEN AND THE DISPLAY WAS OUT OF ELECTRICAL SPECIFICATION WITH MISSING PIXELS. (B)(4).

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471981 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MILACA, INC. 5388

Patients

Seq Age Sex Outcome Treatment
1