FDA Adverse Event Injury Summary report: N

ADAPTA DR

MDR report key: 3991683 · Received August 8, 2014

Report

Report Number
3004209178-2014-14256
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 13, 2014
Report Date
May 13, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NVZ
PMA / PMN Number
P980035
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 507645 LEAD IMPLANTED: (B)(6) 2012. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCE RESULTING FROM A POSSIBLE FRACTURE. THE RV LEAD WAS CAPPED AND REPLACED. DURING THE LEAD REVISION PROCEDURE, IT APPEARED THAT THE PATIENT TWIDDLED THE DEVICE, THEREFORE REMOVING ALL SLACK FROM THE ATRIAL AND VENTRICULAR LEADS. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467831 ADAPTA DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MEDTRONIC MED REL MEDTRONIC PUERTO RICO ADDR01

Patients

Seq Age Sex Outcome Treatment
1 00064 YR Required Intervention 507652 LEAD