FDA Adverse Event
Injury
Summary report: N
ADAPTA DR
MDR report key: 3991683
·
Received August 8, 2014
Report
- Report Number
- 3004209178-2014-14256
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- May 13, 2014
- Report Date
- May 13, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 507645 LEAD IMPLANTED: (B)(6) 2012. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCE RESULTING FROM A POSSIBLE FRACTURE. THE RV LEAD WAS CAPPED AND REPLACED. DURING THE LEAD REVISION PROCEDURE, IT APPEARED THAT THE PATIENT TWIDDLED THE DEVICE, THEREFORE REMOVING ALL SLACK FROM THE ATRIAL AND VENTRICULAR LEADS. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467831 | ADAPTA DR | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | ADDR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00064 YR | Required Intervention | 507652 LEAD |