FDA Adverse Event Injury Summary report: N

SPRINT FIDELIS

MDR report key: 3991665 · Received August 8, 2014

Report

Report Number
2649622-2014-08009
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 14, 2014
Report Date
May 14, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0070-2008
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCTS: (B)(4) ICD 2014 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD INTEGRITY ALERT (LIA) WAS TRIGGERED DUE TO NOISE, A NON-SUSTAINED VENTRICULAR TACHYCARDIA (NSVT) EPISODE AND HIGH IMPEDANCE ON THE LOW VOLTAGE PORTION OF THE RIGHT VENTRICULAR (RV) LEAD. A FRACTURE WAS CONFIRMED. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467686 SPRINT FIDELIS DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694865

Patients

Seq Age Sex Outcome Treatment
1 00070 YR Hospitalization| R 5554-45 LEAD