FDA Adverse Event
Injury
Summary report: N
SPRINT FIDELIS
MDR report key: 3991665
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-08009
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- May 14, 2014
- Report Date
- May 14, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Removal / Correction Number
- Z-0070-2008
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCTS: (B)(4) ICD 2014 (B)(6). (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD INTEGRITY ALERT (LIA) WAS TRIGGERED DUE TO NOISE, A NON-SUSTAINED VENTRICULAR TACHYCARDIA (NSVT) EPISODE AND HIGH IMPEDANCE ON THE LOW VOLTAGE PORTION OF THE RIGHT VENTRICULAR (RV) LEAD. A FRACTURE WAS CONFIRMED. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467686 | SPRINT FIDELIS | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694865 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00070 YR | Hospitalization| R | 5554-45 LEAD |