CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2014-08004
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- May 1, 2014
- Report Date
- May 1, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY : THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL END/ELECTRODES OF THE LEAD WERE BENT, THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A TEAR AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. ELECTRICAL RESISTANCE AND CROSS CONTINUITY TEST RESULTS WERE WITHIN SPECIFICATION. HELIX CANNOT BE EXTEND/ RETRACT DUE TO SENSE ELECTRODE BENT.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THERE WERE MULTIPLE RIGHT VENTRICULAR (RV) LEAD REPOSITION ATTEMPTS AND HELIX EXT ENSION/RETRACTIONS AND THE LEAD IMPEDANCE WAS GREATER THAN 4000 OHMS. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 471116 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5076-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00067 YR |