ADAPTA DR
Report
- Report Number
- 3004209178-2014-14268
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- May 12, 2014
- Report Date
- May 12, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5076-45 LEAD, IMPLANTED: (B)(6) 2006. (B)(4).
PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THAT THE VHR DIAGNOSTICS WERE MISSING/INVALID. THE DIAGNOSTIC MEMORY POINTER CORRUPTED LEADING TO SOME MISSING DATA.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT HAD BEEN IN THE EMERGENCY ROOM WITH PALPITATIONS. WHEN THE DEVICE WAS INTERROGATED A FEW DAY(S) LATER, AN INVALID DATA MESSAGE APPEARED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470879 | ADAPTA DR | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | MEDTRONIC PUERTO RICO OPERATIONS CO. | ADDRL1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00081 YR | 5076-45 LEAD |