FDA Adverse Event Injury Summary report: N

MAXIMO II DR

MDR report key: 3991615 · Received August 8, 2014

Report

Report Number
3004209178-2014-14265
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 1, 2014
Report Date
May 1, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Removal / Correction Number
Z-0113-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN A FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF EITHER THE PRODUCT OR PERFORMANCE DATA, OR COMPLETION OF ANALYSIS, IT COULD NOT BE DETERMINED IF THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 80% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS REPLACED DUE TO END OF SERVICE (EOS) INDICATOR. THE BATTERY LONGEVITY WAS LESS THAN EXPECTED. A NEW DEVICE WAS IMPLANTED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471706 MAXIMO II DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D284DRG

Patients

Seq Age Sex Outcome Treatment
1 00074 YR Hospitalization| R 6947-65 LEAD, 5076-52 LEAD