MAXIMO II DR
Report
- Report Number
- 3004209178-2014-14265
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- May 1, 2014
- Report Date
- May 1, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Removal / Correction Number
- Z-0113-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN A FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF EITHER THE PRODUCT OR PERFORMANCE DATA, OR COMPLETION OF ANALYSIS, IT COULD NOT BE DETERMINED IF THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. (B)(4).
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 80% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE DEVICE WAS REPLACED DUE TO END OF SERVICE (EOS) INDICATOR. THE BATTERY LONGEVITY WAS LESS THAN EXPECTED. A NEW DEVICE WAS IMPLANTED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 471706 | MAXIMO II DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D284DRG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00074 YR | Hospitalization| R | 6947-65 LEAD, 5076-52 LEAD |