FDA Adverse Event
Injury
Summary report: N
CAPSURE Z NOVUS
MDR report key: 3991603
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-08022
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- May 14, 2014
- Report Date
- May 14, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P850089
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT MEDICAL PRODUCTS: A 5054-52 LEAD: (B)(6) 2008. (B)(4).
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE ATRIAL PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE AND OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER).
Description of Event or Problem · 1
IT WAS REPORTED THAT AT A ROUTINE DEVICE CHECK, THE RIGHT ATRIAL (RA) LEAD EXHIBITED AN INCREASE IN IMPEDANCE AND NOISE. A FRACTURE WAS SUSPECTED. THE RA LEAD WAS INACTIVATED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 471631 | CAPSURE Z NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5554-45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00085 YR | Required Intervention | P1501DR IPG |