Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT MEDICAL PRODUCTS: A 5054-52 LEAD: (B)(6) 2008. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE ATRIAL PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE AND OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER).

Description of Event or Problem · 1

IT WAS REPORTED THAT AT A ROUTINE DEVICE CHECK, THE RIGHT ATRIAL (RA) LEAD EXHIBITED AN INCREASE IN IMPEDANCE AND NOISE. A FRACTURE WAS SUSPECTED. THE RA LEAD WAS INACTIVATED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471631 CAPSURE Z NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5554-45

Patients

Seq Age Sex Outcome Treatment
1 00085 YR Required Intervention P1501DR IPG