FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 3991589
·
Received August 8, 2014
Report
- Report Number
- 2182208-2014-02129
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- May 15, 2014
- Report Date
- May 28, 2014
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- KRG
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THAT THE CASE HANDLE WAS BROKEN, THE LATCH TAB WAS BROKEN OFF OF THE POWER CORD BAY DOOR, THE KEYBOARD LATCH WAS BROKEN, THE SYSTEM FAN WAS NOISY AND THE ELECTROCARDIOGRAM (ECG) CONNECTOR WAS LOOSE ON THE LINK ELECTRONIC MODULE (LEM) CIRCUIT BOARD. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THE HANDLE IS BROKEN. THE PROGRAMMER WAS RETURNED FOR REPAIR. IT WAS ANALYZED AND TESTED OUT OF SPECIFICATION. THERE WAS NO PATIENT INVOLVEMENT INDICATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470822 | CARELINK | PROGRAMMER, PACEMAKER | KRG | MEDTRONIC, INC. | 2090W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 2067L RADIOFREQUENCY (RF) HEAD |