SENSIA SR
Report
- Report Number
- 9614453-2014-01690
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 5, 2014
- Report Date
- June 5, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS SENSING INTEGRITY COUNTER (SIC). ANALYST COMMENTED, AUTOMATIC LEAD DIAGNOSTICS SHOWS VENTRICULAR NON-PHYSIOLOGICAL SENSING DETECTED.
IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED AFTER FALLS THAT OCCURRED. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS NOTED TO EXHIBIT LOSS OF CAPTURE AND OVERSENSING WITH POCKET MANIPULATION. DIAGNOSTIC TESTING/TROUBLESHOOTING WAS PERFORMED. THE IPG WAS REPROGRAMMED AND REMAINS IN USE. IN ADDITION IT WAS NOTED THAT A RIGHT VENTRICULAR (RV) LEAD ALERT TRIGGERED FOR LOW PACING IMPEDANCE WITH A NUMBER OF VENTRICULAR TACHYCARDIA (VT) EPISODES THAT APPEARED TO BE NON-PHYSIOLOGICAL. THE RV LEAD WAS REPROGRAMMED AND REMAINS IN USE. HOWEVER, A POSSIBLE BATTERY CHANGE WITH RV LEAD REVISION IS SCHEDULED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470507 | SENSIA SR | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | IPG MFG SWITZERLAND | SESR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R | 5054-58 LEAD |