FDA Adverse Event Injury Summary report: N

SENSIA SR

MDR report key: 3991580 · Received August 8, 2014

Report

Report Number
9614453-2014-01690
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 5, 2014
Report Date
June 5, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS SENSING INTEGRITY COUNTER (SIC). ANALYST COMMENTED, AUTOMATIC LEAD DIAGNOSTICS SHOWS VENTRICULAR NON-PHYSIOLOGICAL SENSING DETECTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED AFTER FALLS THAT OCCURRED. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS NOTED TO EXHIBIT LOSS OF CAPTURE AND OVERSENSING WITH POCKET MANIPULATION. DIAGNOSTIC TESTING/TROUBLESHOOTING WAS PERFORMED. THE IPG WAS REPROGRAMMED AND REMAINS IN USE. IN ADDITION IT WAS NOTED THAT A RIGHT VENTRICULAR (RV) LEAD ALERT TRIGGERED FOR LOW PACING IMPEDANCE WITH A NUMBER OF VENTRICULAR TACHYCARDIA (VT) EPISODES THAT APPEARED TO BE NON-PHYSIOLOGICAL. THE RV LEAD WAS REPROGRAMMED AND REMAINS IN USE. HOWEVER, A POSSIBLE BATTERY CHANGE WITH RV LEAD REVISION IS SCHEDULED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470507 SENSIA SR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND SESR01

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R 5054-58 LEAD