FDA Adverse Event Injury Summary report: N

KAPPA 700 D

MDR report key: 3991525 · Received August 8, 2014

Report

Report Number
9614453-2014-01698
Event Type
Injury
Date Received
August 8, 2014
Date of Event
April 13, 2014
Report Date
May 15, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE U.S. WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS REPLACED DUE TO ELECTIVE REPLACEMENT INDICATOR (ERI). THE BATTERY LONGEVITY WAS LESS THAN EXPECTED. THE IPG WAS REMOVED AND A NEW DEVICE WAS IMPLANTED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470417 KAPPA 700 D PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND KD701

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R