FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3991515 · Received August 8, 2014

Report

Report Number
2649622-2014-08075
Event Type
Injury
Date Received
August 8, 2014
Date of Event
March 20, 2014
Report Date
May 8, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE. CONCOMITANT PRODUCT: 5086MRI45 LEAD, IMPLANTED: (B)(6) 2014. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH POST-OPERATIVE THRESHOLDS DUE TO DISLODGEMENT. THE LEAD WAS REPOSITIONED. HIGHER THRESHOLDS WERE LATER OBSERVED A SECOND TIME DUE TO DISLODGEMENT. THE LEAD WAS REPOSITIONED BUT THE PHYSICIAN WAS NOT SATISFIED WITH THE THRESHOLD MEASUREMENT OBTAINED. THE LEAD WAS ULTIMATELY EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469530 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00075 YR Hospitalization| R A2DR01 IPG