FDA Adverse Event Injury Summary report: N

CAPSURE SP

MDR report key: 3991500 · Received August 8, 2014

Report

Report Number
2649622-2014-08085
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 15, 2014
Report Date
May 15, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P830061
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: SEDR01, IPG, IMPLANTED: (B)(6) 2007. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO ESC (ENVIRONMENTAL STRESS CRACKING) WHILE IN VIVO.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW PACING IMPEDANCE. IN ADDITION, THERE WAS INSULATION DAMAGED NOTED ON THE PROXIMAL END OF THE RIGHT ATRIAL (RA) LEAD. THE RV AND RA LEADS WERE CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469355 CAPSURE SP ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 4524-53

Patients

Seq Age Sex Outcome Treatment
1 00075 YR Hospitalization| R 402458 LEAD