CAPSURE SP
Report
- Report Number
- 2649622-2014-08085
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- May 15, 2014
- Report Date
- May 15, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P830061
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: SEDR01, IPG, IMPLANTED: (B)(6) 2007. (B)(4).
PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO ESC (ENVIRONMENTAL STRESS CRACKING) WHILE IN VIVO.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW PACING IMPEDANCE. IN ADDITION, THERE WAS INSULATION DAMAGED NOTED ON THE PROXIMAL END OF THE RIGHT ATRIAL (RA) LEAD. THE RV AND RA LEADS WERE CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469355 | CAPSURE SP | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 4524-53 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00075 YR | Hospitalization| R | 402458 LEAD |