FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3991481 · Received August 8, 2014

Report

Report Number
2649622-2014-08080
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
March 26, 2014
Report Date
May 12, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: VEDR01 IPG, IMPLANTED: (B)(6) 2009. (B)(4).

Description of Event or Problem · 1

IT WAS LATER REPORTED THAT THE LEAD HAD INFINITE IMPEDANCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD HAD HIGH IMPEDANCE THAT TRIGGERED A LEAD WARNING. SOME OVERSENSING WAS ALSO OBSERVED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471094 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-45

Patients

Seq Age Sex Outcome Treatment
1 00083 YR 5076-58 LEAD