FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 3991443 · Received August 8, 2014

Report

Report Number
2649622-2014-08120
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 15, 2014
Report Date
May 15, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY FOUND NO ANOMALIES.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT PRODUCT: DTBA2D4 ICD, IMPLANTED: (B)(6) 2014; 6935M LEAD, IMPLANTED: (B)(6) 2014. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD DISLODGED, INDUCING VENTRICULAR FIBRILLATION (VF) FOR WHICH THE PATIENT RECEIVED INAPPROPRIATE SHOCKS FROM THE RIGHT VENTRICULAR (RV) LEAD. THE LV LEAD WAS EXPLANTED AND REPLACED AND THE RV LEAD REMAINS IN USE. IT WAS FURTHER REPORTED THAT THE SHOCKS UNEXPECTEDLY REDUCED THE BATTERY LONGEVITY OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WHICH REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470172 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 419678

Patients

Seq Age Sex Outcome Treatment
1 00066 YR Hospitalization| R 4076 LEAD