FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3991441 · Received August 8, 2014

Report

Report Number
2649622-2014-08116
Event Type
Injury
Date Received
August 8, 2014
Date of Event
March 24, 2014
Report Date
May 6, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SHORTLY AFTER IMPLANT THE RIGHT VENTRICULAR (RV) LEAD HAD INCREASED PACING THRESHOLD AND FLUOROSCOPY CONFIRMED THE LEAD WAS DISLODGED. THE LEAD WAS REPOSITIONED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470851 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00076 YR Hospitalization| R 5076-45 LEAD, ADDR01 IPG