FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX
MDR report key: 3991413
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-08128
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- May 9, 2014
- Report Date
- May 9, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P930039
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 4068-52 LEAD IMPLANTED: (B)(6) 1999. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A SUDDEN ONSET OF PECTORAL PACING AND HIGH THRESHOLDS. IT WAS NOTED THAT THE LEAD HAD A POLARITY SWITCH TO UNIPOLAR. LEAD FRACTURE IS SUSPECTED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470675 | CAPSUREFIX | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 4068-58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00067 YR | Hospitalization| R | ADDR01 PACEMAKER |