PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3991378 · Received August 8, 2014

Report

Report Number
2183613-2014-00995
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 27, 2014
Report Date
May 27, 2014
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: ANALYSIS COULD NOT CONFIRM THE REPORTED EVENT, THE DEVICE PASSED FUNCTIONAL TESTING WITH NO ANOMALIES FOUND. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE THE EXTERNAL PULSE GENERATOR (EPG) WAS CONNECTED TO A PATIENT, THE BATTERY WAS REPLACED DURING PACING, HOWEVER THE EPG DID NOT STAY ON DURING THE BATTERY REPLACEMENT AND SHUT DOWN. THE EPG THEN RESTARTED WHEN THE BATTERY WAS REPLACED. THE EPG WAS RETURNED TO THE MANUFACTURER FOR SERVICING. THE PATIENT DID HAVE AN INTRINSIC HEARTBEAT, THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469616 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MILACA, INC. 5388

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention