SPRINT QUATTRO SECURE

MDR report key: 3991351 · Received August 8, 2014

Report

Report Number
2649622-2014-08169
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
April 28, 2014
Report Date
May 5, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. THE ANALYST COMMENTED THAT THE DISTAL END OF THE LEAD DID NOT HAVE A BEND OR OTHERWISE IRREGULAR APPEARANCE THAT WAS OBSERVED DURING ANALYSIS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AT THE IMPLANT PROCEDURE IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A BEND NEAR THE DISTAL COIL UPON REMOVAL FROM THE PACKAGE. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468805 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00071 YR A7700E27 HEART VALVE