FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3991314 · Received August 8, 2014

Report

Report Number
2649622-2014-08193
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 9, 2014
Report Date
June 9, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT OF THE RIGHT VENTRICULAR (RV) LEAD SLOW DEPLOYMENT OF THE HELIX WAS NOTED, TAKING MORE THAN THE EXPECTED TURNS CAUSING PLACEMENT DIFFICULTY. THE FOLLOWING DAY IT WAS NOTED THAT THE RV LEAD HAD DISLODGED, HIGH THRESHOLDS AND HIGH IMPEDANCE WERE FOUND. SURGICAL REPOSITIONING WAS SUCCESSFUL, AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472090 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00062 YR Hospitalization| R 5086MRI45 LEAD, A2DR01 IPG