FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX MRI SURESCAN
MDR report key: 3991314
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-08193
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 9, 2014
- Report Date
- June 9, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT OF THE RIGHT VENTRICULAR (RV) LEAD SLOW DEPLOYMENT OF THE HELIX WAS NOTED, TAKING MORE THAN THE EXPECTED TURNS CAUSING PLACEMENT DIFFICULTY. THE FOLLOWING DAY IT WAS NOTED THAT THE RV LEAD HAD DISLODGED, HIGH THRESHOLDS AND HIGH IMPEDANCE WERE FOUND. SURGICAL REPOSITIONING WAS SUCCESSFUL, AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 472090 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00062 YR | Hospitalization| R | 5086MRI45 LEAD, A2DR01 IPG |