FDA Adverse Event Malfunction Summary report: N

ATTAIN OTW

MDR report key: 3991283 · Received August 8, 2014

Report

Report Number
2649622-2014-08179
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 19, 2014
Report Date
May 19, 2014
Manufacturer
MPRI
Product Code
LWP
PMA / PMN Number
P010015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: MODEL: 6947, LEAD, IMPLANT: (B)(6) 2008; MODEL: 5076-52, LEAD, IMPLANT: (B)(6) 2007. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO. THE OUTER INSULATION ALSO DEVELOPED A BREACH DUE TO ESC (ENVIRONMENTAL STRESS CRACKING) WHILE IN VIVO. A COSMETIC DEPRESSION AND COSMETIC ESC (ENVIRONMENTAL STRESS CRACKING) WERE ALSO NOTED. THE OUTER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION.

Description of Event or Problem · 1

IT WAS REPORTED THAT A LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLDS. THE PHYSICIAN CHOSE TO REMOVE AND REPLACE THE LV LEAD AT THE TIME OF THE GENERATOR CHANGE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472802 ATTAIN OTW PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MPRI 419488

Patients

Seq Age Sex Outcome Treatment
1 00082 YR D224TRK, ICD