FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3991278 · Received August 8, 2014

Report

Report Number
2649622-2014-08191
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 9, 2014
Report Date
May 9, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. THE ANALYST COMMENTED, THE LEAD WAS RETURNED WITH THE HELIX BENT AND BLOOD IN THE SLEEVEHEAD. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, WHILE TRYING TO SCREW THE LEAD INTO THE RIGHT VENTRICULAR APEX, THE HELIX DID NOT DEPLOY. IT WAS NOTED THAT THE HELIX WAS STUCK. ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467961 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00063 YR