FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 3991268 · Received August 8, 2014

Report

Report Number
2649622-2014-08221
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 14, 2014
Report Date
May 14, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - A PROXIMAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: (B)(4) ICD, IMPLANTED: 2010 (B)(6); 6947 LEAD, IMPLANTED 2010 (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED DIAPHRAGMATIC NERVE STIMULATION. THE LEFT VENTRICULAR (LV) LEAD WAS PULLED BACK AND THE TIP WAS IN THE SUPERIOR VENA CAVA. THEREFORE, THE LEAD COULD NOT BE TOTALLY REMOVED AND WAS CUT AND CAPPED. A PORTION WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472798 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 419688

Patients

Seq Age Sex Outcome Treatment
1 00056 YR Hospitalization| R 5076 LEAD