FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3991255 · Received August 8, 2014

Report

Report Number
2649622-2014-08231
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT: 429678 LEAD IMPLANTED: (B)(6) 2013; 5076-45 LEAD IMPLANTED: (B)(6) 2003. (B)(4). EVALUATION SUMMARY: THE RV (RIGHT VENTRICULAR) DEFIBRILLATION COIL DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO; FULL LEAD RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A POSSIBLE LEAD FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468483 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00089 YR Hospitalization| R DTBA1D4 CRT-D