FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3991249 · Received August 8, 2014

Report

Report Number
2649622-2014-08228
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 6, 2014
Report Date
October 24, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SAVE TO DISK CLINICAL DATA REVIEW WAS DETERMINED TO BE NOT REQUIRED BECAUSE PHYSICAL ANALYSIS WAS COMPLETED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD HIGH IMPEDANCE AND FRACTURED. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468481 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R