FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3991227 · Received August 8, 2014

Report

Report Number
2649622-2014-08240
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 17, 2014
Report Date
June 17, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR OF THE LEAD WAS EXTRINSICALLY OVER-ROTATED. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. THE LEAD WAS RECEIVED WITH THE HELIX FULLY EXTENDED, AND DISTORTION OF THE DISTAL CONDUCTOR WITHIN THE IS-1 CONNECTOR. THIS IS INDICATIVE OF THE CONNECTOR PIN BEING TURNED AN EXCESSIVE NUMBER OF TIMES DURING HELIX EXTENSION.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE RIGHT VENTRICULAR (RV) LEAD HELIX WOULD NOT RETRACT. THE LEAD WAS ATTEMPTED AND NOT USED. ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472567 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00069 YR